Diados Diagnostic Detector

Model/Serial: 60004-3356 (purchased: 2018)

Description: X-ray-semiconductor (40 - 150 kV) for use with Diados.

PTW Calibration: 09-Mar-2018

Seibersdorf Diagnostic Quality Filter Range Na ± U [Gy ± Gy]
08.06.2022 RQR5 (DV70) 2.5 mm Al Dose Rate Low 0.992 ± 0.018
Dose Rate Medium 0.992 ± 0.018
Dose Rate High 0.992 ± 0.018
Dose Rate Auto 0.995 ± 0.018
Dose Low 0.991 ± 0.018
Dose Medium 0.992 ± 0.018
Dose High 0.992 ± 0.018
Dosis Auto 0.992 ± 0.018
RQA5 (DN70) 23.5 mm Al Dosis Auto 0.992 ± 0.018

Note: U = k uc is the expanded uncertainty with the coverage factor k = 2.

Logbook

Date Event
28.03.2018 Delivery of detector
11.06.2018 Seibersdorf calibration (Na ~ 1.000)
08.06.2022 Seivrsdorf verification and calibration

back to equipment