QUICKCHECKwebline

Model/Serial: T42031 #1933
Compensator 6FFF: T42031.1.310-120531
Compensator 10FFF: T42031.1.320-120532

Model/Serial: T42031 #1918
Compensator 6FFF: T42031.1.310-120536
Compensator 10FFF: T42031.1.320-120536

Description: Morning check device with 13 ionization chambers.

Accessory: Holder T40003.1.003 for Varian Accessory Mount 889300-03.

Measurement Chain: QUICKCHECKwebline > Docking Station > LAN > PC

Images

QUICKCHECKwebline with 10FFF compensator on the TrueBeamSTx:

QUICKCHECKwebline with 10FFF compensator on the TrueBeamSTx

Check of dosimetric equivalence of the two compensator sets (CT scan 120 kV):

Compensator comparison

Logbook

 
 
Date Event
26.07.2012 Delivery (new device) of T42031 #557.
20.12.2012 Delivery (new device) of T42031 #621.
26.07.2013 #0557: Sent to PTW for repair ("measures without radiation = auto-trigger").
21.08.2013 #0557: Back from repair.
25.09.2014 #0621: Sent to PTW for repair ("displayed temperature: -54.6°C").
30.10.2014 #0621: Back from repair.
03.11.2014 #0557: Sent to PTW for repair ("displayed temperature: -54.6°C").
04.12.2014 #0557: Back from repair.
08.08.2016 #0942: Received loan unit according to service contract.
10.08.2016 #0621: Sent to PTW (displayed temperature: -54.6°C, pressure off by 4 hPa, auto-trigger).
25.08.2016 #1918: Recieved new device from PTW (replaces #621).
26.08.2016 #0557: Sent to PTW for repair (similar problems as #621).
30.09.2016 #1933: Recieved new device from PTW (replaces #557).
06.10.2016 #0942: Returned loan unit.
19.10.2017 #0998: Received loan unit.
23.10.2017 #1933: Sent to PTW for repair (auto-trigger).
22.11.2017 #1933: Back from repair.
28.11.2017 #0998: Returned loan unit.
19.02.2018 #1505: Received loan unit.
20.02.2018 #1918: Sent to PTW for repair (date/time has to be set on every use).
26.03.2018 #1918: Back from repair.
28.03.2018 #1505: Returned loan unit.
11.05.2018 #2485: Received loan unit.
15.05.2018 #1933: Sent to PTW for repair (date/time has to be set on every use).
10.07.2018 #1933: Back from repair.
15.07.2018 #2485: Returned loan unit.
18.10.2019 #1918: Date/time has to be set on every use. Requested loan unit from PTW.
24.10.2019 #0942: Received loan unit.
28.10.2019 #1918: Sent to PTW for repair.
21.11.2019 #1918: Back from repair.
25.11.2019 #0942: Returned loan unit.
17.02.2020 #1933: Date/time has to be set on every use. Requested loan unit from PTW.
20.02.2020 #0807: Received loan unit.
24.02.2020 #1933: Sent to PTW for repair.
18.03.2020 #1933: Back from repair.
18.03.2020 #0807: Returned loan unit.
04.05.2021 #1918: Pressure display not stable. Requested loan unit from PTW.
10.05.2021 #2603: Received loan unit.
11.05.2021 #1918: Sent to PTW for repair.
22.06.2021 #1918: Back from repair.
28.06.2021 #2603: Returned loan unit.
26.07.2021 #1933: Starts measurements without radiation. Requested loan unit from PTW.
02.08.2021 #2038: Received loan unit.
06.08.2021 #1933: Sent to PTW for repair.
24.08.2021 #1933: Back from repair.
31.08.2021 #2038: Returned loan unit.
16.02.2023 #1918 and #1933: Date/Time cannot be saved. Requested two loan units from PTW.
22.02.2023 #2038 and #3281: Received loan units.
27.02.2023 #1918 and #1933: Sent to PTW for repair.
31.03.2023 #1918: Back from repair (clock module defective).
05.04.2023 #1933: Back from repair (clock module defective).
13.04.2023 #2038 and #3281: Returned loan units.

back to equipment

Valid XHTML 1.0 Strict